The Membrane Sensors Sensitive to Fluoxetine - Optimization of the Analytical Parameters
PDF

Keywords

 Ion-selective electrode, potentiometric titration, drug analysis, fluoxetine hydrochloride.

How to Cite

Joanna Drozd. (2015). The Membrane Sensors Sensitive to Fluoxetine - Optimization of the Analytical Parameters. Journal of Basic & Applied Sciences, 11, 397–402. https://doi.org/10.6000/1927-5129.2015.11.56

Abstract

Background: The analytical characteristics of electrodes containing liquid PVC membrane depend on the purity of the electroactive material and the applied membrane solvents.
Methods: Fluoxetine - one H-bond donor, formed ion-associated complexes with tetraphenylborate, tetrakis(4-chlorophenyl)borate and dipicrylamine. To confirm the quality and purity obtained complexes the elemental and TLC analysis were performed. The construction and general characteristics of fluoxetine ion-selective plastic membrane sensors, based on the use of a fluoxetine – tetrakis(4-chlorophenyl)borate ion-pair complex as electroactive material with 2-nitrophenyloctyl ether, bis(2-ethylhexyl)sebacate, bis(2-butylpentyl)adipate and 1-isopropyl-4-nitrobenzene as solvent mediators.
Results: The electrode containing Fl-ClTPB-NPhOE (fluoxetine-tetrakis(4-chlorophenyl)borate-2-nitrophenyloctyl ether showed a linear response to fluoxetine at concentration ranges of 30.93 – 2.16 g·L-1 with a super Nernstian cationic slope 60.19 mV·decade-1.
Conclusions: This Fl-ClTPB-NPhOE sensor was used for the potentiometric determination of the content of fluoxetine hydrochloride in pure form and pharmaceutical formula.

https://doi.org/10.6000/1927-5129.2015.11.56
PDF

References

Eugster R, Rosatzin T, Rusterholz B, et al. Plasticisers for liquid polymeric membranes of ion-selective chemical sensors. Anal Chim Acta 1994; 289: 1-13. http://dx.doi.org/10.1016/0003-2670(94)80001-4

Ganjali MR, Haji–Hashemi H, Faridbod F, Norouzi P, Qomi M. Potentiometric determination of alprazolam based on carbon paste and PVC membrane electrodes. Int J Electrochem Sci 2012; 7: 1470-1481.

Ezzeldin E, Hefnawy MM, Abounassif MA, Tammam MH, Mostafaet GAE. PVC membrane sensors for potentiometric determination of bambuterol in pharmaceutical formulation. Int J Electrochem Sci 2012; 7: 10570-10581.

Al-Mustafa JI, Abu-Dalo MA, Nassory NS. Liquid selective electrodes for dextromethorphan hydrobromide based on a molecularly imprinted polymer in PVC matrix membrane. Int J Electrochem Sci 2014; 9: 292-303.

Eman YZF, Tamer AA, Gehad GM, Yusra HHA. Construction of different types of ion-selective electrodes. Characteristic performances and validation for direct potentiometric determination of orphenadrine citrate. Int J Electrochem Sci 2012; 7: 4443-4464.

Gupta VK, Agarwal S, Singhal B. Potentiometric assay of antipsychotic drug (ziprasidone hydrochloride) in pharmaceuticals, serum and urine. Int J Electrochem Sci 2011; 6: 3036-3056.

Malongo TK, Blankert B, Kambu O, Amighi K, Nsangu J, Kauffmannet JM. Amodiaquine polymeric membrane electrode. J Pharmaceut Biomed Anal 2006; 41: 70-76. http://dx.doi.org/10.1016/j.jpba.2005.10.014

Ganjali MR, Karimi A, Norouzi P. Letrozole potentiometric PVC membrane and nano-composite carbon paste electrodes. Int J Electrochem Sci 2012; 7: 3681-3692.

Ganjali MR, Larijani B, Norouzi P. Determination of midazolam by potentiometric PVC membrane and MWCNTS based carbon paste sensors. Int J Electrochem Sci 2012; 7: 4822-4833.

Ganjali MR, Aghili S, Hamedani MP, Norouzi P. Potentiometric determination of raloxifene by PVC membrane sensor. Int J Electrochem Sci 2012; 7: 10394-10403.

Ganjali MR, Karimi A, Shahtaheri SJ, Norouziet P. Determination of clonidine by potentiometry using PVC membrane electrode. Int J Electrochem Sci 2013; 8: 1999-2008.

Faridbod F, Khamseh-Nejad M, Ganjali MR, Norouzi P, Hajiaghababaei L. Cetrimide potentiometric PVC membrane sensor. Int J Electrochem Sci 2012; 7: 1917-1926.

Faridbod F, Hosseini M, Ganjali MR, Norouzi P. Potentiometric sensor for determination of clomiphene. Int J Electrochem Sci 2013; 8: 1976-1985.

Faridbod F, Mizani F, Ganjali MR, Norouzi P. Determination of ondansetron hydrochloride by a liquid membrane potentiometric sensor based on room temperature ionic liquids. Int J Electrochem Sci 2013; 8: 10461-10472.

Faridbod F, Ganjali MR, Norouzi P. Potentiometric PVC membrane sensor for the determination of terbinafine. Int J Electrochem Sci 2013; 8: 6107-6117.

Ibrahim H, Issa YM, Abu-Shawish HM. Potentiometric flow injection analysis of dicyclomine hydrochloride in serum, urine and milk. Anal Chim Acta 2005; 532: 79-88.

Nassory NS, Maki SA, Al-Phalahy BA. Preparation and potentiometric study of promethazine hydrochloride selective electrodes and their use in determining some drugs. Turk J Chem 2008; 32: 539-548.

Hahassouna MEM, Elsuccary SAA. PVC membrane electrode for the potentiometric determination of ipratropium bromide using batch and flow injection techniques. Talanta 2008; 75: 1175-1183. http://dx.doi.org/10.1016/j.talanta.2008.01.014

Sales MGF, Tomás JFC, Lavandeira SR. Flow injection potentiometric determination of chlorpromazine. J Pharmaceut Biomed Anal 2006; 41: 1280-1286. http://dx.doi.org/10.1016/j.jpba.2006.03.027

Drozd J, Gumieniczek A. Meclophenoxate ion-selective plastic membrane sensors and their pharmaceutical applications. Curr Issues Pharm Med Sci 2012; 25: 7-11. http://dx.doi.org/10.12923/j.2084-980X/25.3/a.24

Riad SM, Mostafa NM. Ion selective electrodes for potentiometric determination of baclofen in pharmaceutical preparations. Anal Bioanal Electrochem 2013; 5: 494-505.

Hussien EM, Abdel-Gawad FM, Issa YM. Ion-selective electrodes for determination of fluoxetine in capsules and in biological fluids. Biochem Eng J 2011; 53: 210-215. http://dx.doi.org/10.1016/j.bej.2010.10.015

Ganjali MR, Faridbod F, Rashedi H, Hosseini M, Hosseini M, Norouzet P. Fluoxetine determination by PVC membrane and nano-composite carbon paste electrodes. Int J Electrochem Sci 2011; 6: 2299-2311.

Craggs A, Moody GJ, Thomas JSDR. PVC matrix membrane ion-selective electrodes: Construction and laboratory experiments. J Chem Educ 1974; 51: 541-544. http://dx.doi.org/10.1021/ed051p541

Siska E, Pungor E. Potentiometric determination of tetraphenylborate ions with silver nitrate. Determination of silver, potassium and tallium (I). Fresenius’ Zeitschriftfür Analytische Chemie 1971; 257: 12-18. http://dx.doi.org/10.1007/BF00424177

Misztal G, Skibi?ski R. Chromatographic analysis of new antidepressant drugs by normal - and reversed-phase TLC. J Planar Chromatogr – Modern TLC 2001; 14: 300-304. http://dx.doi.org/10.1556/JPC.14.2001.4.14

Pungor E, Toth K, Hrabeczy-Pall A. Selectivity coefficients of ion-selective electrode. Pure Appl Chem 1979; 51: 1913-1980.

Mascini M. Uses of known addition Gran’s plots and the related methods with ion-selective electrodes. Ion-Selective Electrodes Reviews 1980; 2: 17-21.

Creative Commons License

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

Copyright (c) 2015 Joanna Drozd